FSP065M-DUA – 64W 90-264 VAC medical desktop power supply
- Technology
- Desktop power supplies
- Partner
- FSP Protek
The FSP065M-DUA medical USB PD power supply is an external 65W AC/DC adapter for medical equipment using a USB Type-C power interface. It accepts a 90–264 VAC input and uses an IEC 320/C14 inlet, allowing the mains lead to be selected separately for the destination market. Five USB Power Delivery profiles provide 5V, 9V, 12V, 15V or 20V, with a maximum output of 3.25A at 20V. Its Class I design incorporates protective earth and provides 2 MOPP isolation between the input and output. Controlled touch and earth leakage currents support integration into equipment subject to medical electrical safety requirements. DOE Level VI efficiency, no-load consumption of up to 0.21W and EN 55011 Class B emissions address energy and EMC considerations. The unit is intended for engineers developing medical equipment that requires an external, multi-voltage USB-C PD supply.

The FSP065M-DUA is a medical-grade external desktop power supply designed for equipment that uses a USB Type-C Power Delivery input. It combines a universal 90–264 VAC input range with an IEC 320/C14 inlet, making it suitable for international deployment with a region-specific mains lead.
With five USB PD output profiles from 5V to 20V and up to 65W maximum output power, it supports a wide range of medical device power architectures. Class I construction, 2 MOPP input-to-output isolation, controlled leakage currents and IEC 60601-1 related compliance characteristics make it suitable for integration into professional medical equipment designs.
Range features
A high level overview of what this range offers
- 65W USB Type-C Power Delivery output – Provides one power interface for compatible equipment operating from 5V to 20V.
- Five selectable voltage profiles – Supports 5V, 9V, 12V, 15V and 20V power architectures.
- 90–264 VAC input range – Enables operation across common regional mains supplies.
- IEC 320/C14 inlet – Allows a detachable, region-specific mains cable to be used.
- IEC 60601-1 medical safety compliance – Supports integration into medical electrical equipment designs.
- Class I construction with protective earth – Provides an earthed external power architecture.
- 4,000 VAC input-to-output withstand voltage – Provides 2 MOPP isolation across the primary-to-secondary barrier.
- Touch current up to 100µA – Helps control accessible leakage current at the specified test condition.
- DOE Level VI efficiency – Addresses external power supply energy-efficiency requirements.
- No-load consumption up to 0.21W – Limits standby power when no output load is connected.
- EN 55011 Class B conducted and radiated emissions – Supports EMC planning for medical equipment.
- OVP, OCP, short-circuit and OTP protection – Provides defined responses to common electrical and thermal fault conditions.
- 5,000m operating altitude – Supports deployment in high-altitude locations within the stated environmental limits.
- 1.5m fixed USB Type-C cable – Provides a ready-to-connect DC output without a separate output lead.
Downloads
for FSP065M-DUA - 64W 90-264 VAC medical desktop power supply
What’s in this range?
All the variants in the range and a comparison of what they offer
| Specification | FSP065M-DUA |
|---|---|
Product type | Medical USB PD desktop power supply |
Maximum continuous output power | 65W at 40°C operating temperature |
AC inlet | IEC 320/C14 |
Input voltage | 90–264 VAC |
Input frequency | 47–63Hz |
Input current | ≤1.8A RMS at 100 VAC; ≤1.0A RMS at 240 VAC |
Input protection | T3.15AH/250V internal fuses fitted in line and neutral |
Construction | Class I with protective earth |
Output connector | USB Type-C |
Output-voltage profiles | 5V / 9V / 12V / 15V / 20V |
Minimum output current | 0A |
Maximum output current | 3A / 3A / 3A / 3A / 3.25A |
Maximum power by profile | 15W at 5V; 27W at 9V; 36W at 12V; 45W at 15V; 65W at 20V |
Output-voltage tolerance | ±5% |
Ripple and noise | ≤250mV at 5V, 9V and 12V; ≤300mV at 15V and 20V |
Average active efficiency at 115 VAC | 85.72% at 5V/2A; 84.84% at 5V/3A; 88.55% at 9V; 89.60% at 12V; 90.39% at 15V; 90.27% at 20V |
Average active efficiency at 230 VAC | 82.31% at 5V/2A; 81.81% at 5V/3A; 86.95% at 9V; 88.62% at 12V; 90.05% at 15V; 91.13% at 20V |
Energy-efficiency level | DOE Level VI |
No-load power consumption | ≤0.21W |
Hold-up time | ≥6ms at 100 VAC/60Hz or 230 VAC/50Hz |
Line regulation | ±1% maximum at full load |
Touch current | ≤100µA at 264 VAC, 63Hz |
Earth leakage current | ≤150µA at 264 VAC, 63Hz |
Input-to-output withstand voltage | 4,000 VAC, 2 MOPP |
Input-to-ground withstand voltage | 1,500 VAC, 1 MOPP |
Output-to-ground withstand voltage | 500 VAC |
Overvoltage protection | Latch-off |
Overcurrent and short-circuit protection | Automatic recovery |
Overtemperature protection | Latch-off |
Conducted and radiated emissions | EN 55011 Class B |
Electrostatic discharge immunity | ±15kV air; ±8kV contact |
Radiated immunity | 3V/m |
Fast transient/burst immunity | ±2kV |
Surge immunity | ±1kV differential mode; ±2kV common mode |
Conducted immunity | 3Vrms |
Magnetic-field immunity | 30A/m |
Operating temperature | 0°C to +40°C |
Storage temperature | −20°C to +85°C |
Operating humidity | 5% to 95% RH, non-condensing |
Storage humidity | 5% to 95% RH, non-condensing |
Maximum operating altitude | 5,000m |
MTBF | 358,302 hours at full load and 25°C ambient, calculated in accordance with SR332 |
Case dimensions | 114 ±0.3 × 50 ±0.3 × 34 ±0.3mm |
Output cable length | 1,500 ±20mm |
Approximate weight | 194.5g |
Case colour | Black |
Output cable conductors | VBUS and GND: 20AWG; CC: 26AWG |
Principal safety standards | IEC 60601-1 and IEC 62368-1 |
Medical EMC standard | IEC 60601-1-2 |
FAQs
for FSP065M-DUA – 64W 90-264 VAC medical desktop power supply
The rated maximum output is 65W when the 20V USB PD profile is operating at up to 3.25A. The product is therefore a 65W supply rather than a 64W model.
No. The available maximum power depends on the negotiated USB PD profile: 15W at 5V, 27W at 9V, 36W at 12V, 45W at 15V and 65W at 20V.
It operates from 90–264 VAC at 47–63Hz. Input current is specified as no more than 1.8A RMS at 100 VAC and 1.0A RMS at 240 VAC.
The FSP065M-DUA uses an IEC 320/C14 appliance inlet. A suitable three-conductor mains cable with the required regional plug must be selected separately.
The input-to-output barrier is rated at 4,000 VAC and 2 MOPP. The input-to-ground rating is 1,500 VAC at 1 MOPP, while the output-to-ground withstand rating is 500 VAC.
Overvoltage and overtemperature conditions use a latch-off response. Overcurrent and short-circuit conditions use automatic recovery after the fault is removed.
Touch current is limited to 100µA and earth leakage current to 150µA, both measured at 264 VAC and 63Hz. These values should be considered as part of the complete equipment leakage-current assessment.
The operating range is 0°C to +40°C with 5% to 95% relative humidity, non-condensing. The storage range is −20°C to +85°C, and operation is specified at altitudes up to 5,000m.
The adapter body measures approximately 114 × 50 × 34mm and weighs about 194.5g. It has a fixed USB Type-C output cable measuring approximately 1.5m.
No. The power supply approvals can support the equipment certification process, but the completed medical device must still be assessed against the standards and requirements applicable to the full system and its intended use.